刘吉强 李留斌 周承强 朱建林 延玲 杨晗玉



摘 要:介绍了INFICON 3000 微型气相色谱仪在炼厂有限空间爆炸气检测中的应用,通过对微型气相色谱柱温、载气压力的实验条件进行讨论,确定了最佳的方法设置:通道A、B柱温为60 ℃,通道C柱温为90 ℃,载气压力通道A、B、C均设置为30 psi,该方法的相对误差和相对标准偏差均小于2%,样品全组分分析仅需2 min。与常规气相色谱SP 3420A型气相色谱仪相比,相对偏差在20%以内,且其灵敏度更高,更能反映样品的真实性。
关 键 词:微型气相色谱; 常规气相色谱; 有限空间; 爆炸气
中图分类号:O 657 文献标识码: A 文章编号: 1671-0460(2016)04-0832-03
Abstract: A micro-gas chromatography was introduced into determination of explosive gas in confined space. The instrument setting of column temperature and carrier gas pressure of micro-gas chromatography were discussed. The results show that the optimized column temperature of channel A and B is 60 ℃,that of channel C is 90 ℃.The optimized carrier gas pressure of the three channels is 30 psi.The relative deviation and relative standard deviation are all smaller than 2% ,and the total time of the running cycle is only 2 min for a full analysis. Compared with normal gas chromatography SP3420A, the relative deviation is within 20% ,and the micro-gas chromatography is more sensitive.
Key words: Micro-gas chromatography; Normal gas chromatography; Confined space; Explosive gas
爆炸气分析是炼油化工装置塔、罐、容器等空间有限装置开停工检修最基本的、必要的安全检验,是涉及能否动工检修和人身、设备安全不可缺少的步骤,而检测有限空间中爆炸气体、有毒气体及氧气的含量是判断许可作业的重要依据之一[1-3]。
目前普遍采用常规气相色谱仪来实现对有限空间爆炸气的检测[4]。虽然用常规气相色谱法分析爆炸气样品,完全可以满足对爆炸气样品中各组分分析结果的准确性和可靠性的分析要求,但是,常规气相色谱仪具有分析速度慢的特点,分析一个爆炸气样品的分析耗时至少在15 min以上,全组分分析耗时至少在20 min以上。根据有限空间作业安全管理规定,有限空间中作业前应进行气体检测,注明检测时间和结果,在结果报出30 min后,仍未开始作业,应重新进行检测。如作业中断,再进入之前应重新进行气体检测[5]。……