XN-9000全血细胞分析仪性能评价
兰小英,黄 宇,苏秋妮
(厦门大学附属第一医院,福建厦门 361003)
目的:对Sysmex公司的XN-9000全血细胞分析仪进行性能验证,以便为临床提供更好的质量保证。方法:按照中华人民共和国卫生行业标准《临床血液学检验常规项目分析质量要求》(WS/T406-2012)对XN-9000进行精密度,准确度,可报告范围,携带污染率的验证。结果:该仪器的精密度,准确度,可报告范围,携带污染率均在允许范围内。结论:XN-9000各方面性能良好,测定结果准确,可为临床疾病的诊断和治疗提供依据。
性能验证;精密度;准确度;可报告范围;携带污染率
[ABSTRACT]Objective:To evaluate the complete blood count performance quality of Sysmex XN-9000 Automatic Hematology Analyzer to provide better clinical quality assurance.Methods:The precision,accuracy,reportable range and residual contamination rate of XN-9000 Automatic Hematology Analyzer were investigated in accordance with"Test Routine Clinical Hematology Analysis Project Quality Requirement(WS/T406-2012)"ofthe People's Republic of China Health Industry Standard.Results:The precision,accuracy,reportable range and residual contamination rate of XN-9000 Automatic Hematology Analyzer were respectively within acceptable limits.Conclusion:The performance and accurate results of XN-9000 Automatic Hematology Analyzer in Complete Blood Count are overall excellent and in good condition ,providing the reliable clinical data.
[KEYWORDS]performance evaluation;precision;accuracy;reportable range;residual contamination rate
随着科技的发展,精密、准确、快速的自动化全血细胞分析仪应运而生。为了解Sysmex公司的XN-9000全血细胞分析仪的工作性能,本研究对其精密度、准确度、可报告范围、携带污染率进行评价,现报告如下。
1.1 标本收集
门诊患者新鲜EDTA-K2抗凝全血,充分混匀及时送检,2 h内检测完毕。
1.2 仪器与试剂
配套的Sysmex XN-9000全血细胞分析仪校准品、质控品和试剂。
1.3 方法
性能评价前先定标,定标是用机器配套的标准品,然后每天做质控,在仪器状态良好的情况下,按照上述标准的要求对XN-9000分析仪检测白细胞(WBC)红细胞(RBC)血红蛋白(HGB)血细胞压积(HCT)平均红细胞容积(MCV)和血小板(PLT)的精密度、准确度、可报告范围、携带污染率进行性能评价。
1.3.1 精密度 批内精密度:根据NCCLS EP15-A文件,取低、高两个水平新鲜全血标本,连续重复测定10次,计算批内变异系数(CV批内)。……