化学需氧量快消比色皿法低量程实验条件探讨

2017-09-09 09:25:54廖成朋赵汝汝
科技资讯 2017年22期

廖成朋++赵汝汝

DOI:10.16661/j.cnki.1672-3791.2017.22.210

摘 要:对水质中化学需氧量测定的快速消解分光光度法(HJ/T 399—2007)中低量程实验条件进行探讨,相对对误差在0.00%~1.20%之间,精密度(RSD)为0.00%~0.74%。并用重铬酸盐法和本法对低浓度水样进行对比分析,证明采用两种方法测定结果无显著差异。

关键词:化学需氧量 快速消解分光光度法 重铬酸盐法 低量程实验

中图分类号:X82 文献标识码:A 文章编号:1672-3791(2017)08(a)-0210-04

Discussion on Low Range Experimental Conditions of Chemical Oxygen Demand Rapid Colorimetric Metho

Liao Chengpeng1 Zhao Ruru2

(1.Chongqing state environmental monitoring Co Ltd; 2.Technical center of Chongqing entry exit inspection and Quarantine Bureau, Chongqing, 400020, China)

Abstract:The low range experimental conditions of rapid digestion spectrophotometry (HJ/T 399 - 2007) for determination of COD in water quality are discussed. The relative error is between 0.00%~1.20% and the precision (RSD) is 0.00%~0.74%. The comparison of dichromate method with this method in low concentration water samples showed that there was no significant difference between the two methods.

Key Words: Chemical oxygen demand; Rapid digestion spectrophotometry; Dichromate process; Low range experiment

化學需氧量是指标水体有机污染的一项重要指标,能够反应出水体的污染程度,数值越大表明水体的污染情况越严重[1]。快速消解分光光度法[2]与重铬酸盐法[3]相比耗时短、能耗低、试剂用量少,能有效减少银盐、汞盐、铬盐带来的二次污染,适用于日常及应急监测。近年来,人们对不同实验条件进行研究[4-6]和方法对比[7-11],并未对最佳测定波长及试剂中沉淀干扰进行探讨。

当消解管作为比色管比色时会受到较多影响,如管壁的均匀性、透光性、划痕印记等,因此选用比色皿法可以排除以上因素带来的误差。文献[6]指出试剂中沉淀影响吸光度测量,并未对其解决方法进行探讨,在绘制低量程工作曲线时,由于过量的硫酸汞在酸性溶液里存在电离平衡:HgSO4=Hg2++SO42-①,加入硫酸银后:Ag2SO4=2Ag++SO42-②,硫酸根离子的增加使反应式①平衡向左移动,生成硫酸汞,这部分物质得不到及时溶解,就沉淀析出;当加入3.00 mL试样时,由于硫酸汞和硫酸银微溶于水,水样的的稀释作用使其溶解度迅速减小而导致析出。但此时被稀释的硫酸粘度较高,沉淀难以沉降聚集在消解管底部,影响吸光度测量。……

登录APP查看全文