原子吸收分光光度计线性误差测量结果的不确定度评定

2018-04-21 13:31:56陶成綦心滕飞
品牌与标准化 2018年4期

陶成 綦心 滕飞

【摘要】 根据JJG 694-2009《原子吸收分光光度计》中关于原子吸收分光光度计线性误差的检定方法,分析了测量过程中引入的不确定度分量,对原子吸收分光光度计线性误差的测量结果进行不确定度评定。

【关键词】 线性误差;原子吸收分光光度计;不确定度

【DOI编码】 10.3969/j.issn.1674-4977.2018.04.006

Evaluation of Uncertainty in Measurement Results of the Linear Error of the Atomic Absorption Spectrophotometer

TAO Cheng,QI Xin,TENG Fei

(Liaoning Provincial Institute of Measurement,Shenyang 110004,China)

Abstract: According to the calibration method of the linear error of the atomic absorption spectrophotometer in the JJG 694-2009 Atomic Absorption Spectrophotometers,the uncertainty components introduced in the measurement process were analyzed,and the uncertainty of the measurement results of the linear error of the atomic absorption spectrophotometer were evaluated.

Key words: linear error;atomic absorption spectrophotometer;uncertainty

原子吸收分光光度计是根据被测元素的基态原子对特征辐射的吸收程度进行定量分析的仪器,其测量原理基于朗博比尔光吸收定律。按照原子化器类型,仪器可以分为火焰原子化器和非火焰(石墨炉法)原子化器。

1 概述

1.1 测量依据

JJG 694-2009《原子吸收分光光度计》。

1.2 环境条件

温度10~35℃;湿度≤85%。

1.3 计量标准

主要计量标准设备为原子吸收分光光度计检定用标准物质,标准物质证书号:GBW(E)130079。具体数据见表1。

1.4 校准对象

校准的原子吸收分光光度计按照原子化原理可以分为火焰法和石墨炉法。火焰法使用铜标准溶液测定,石墨炉法使用镉标准溶液测定,本次检测的对象是一台由德国耶拿公司生产的ZEEnit700原子吸收分光光度计。

1.5 测量方法

根据规程要求设定仪器参数,火焰法和石墨炉法分别测量Cu标准溶液和Cd标准溶液,对曲线的每一点重复测量3次,取3次测量平均值,按线性回归法绘制标准曲线,计算曲线中间点的线性误差。

2 建立数学模型,列出不确定度传播率

2.1 数学模型

按照JJG 694-2009中要求,选择0.0,0.5,1.0,3.0?g/mL标准溶液系列,测定仪器的线性误差,线性误差的数学模型为:

式中:[Δxi]——线性误差,%;

[xi]——第i点按照线性方程计算出的测得浓度值,?g/mL;