王帅彬



摘 要
LB770低本底α/β计数器定期标定维护,在进行本底测量时发现1/2/6道本底偏大,最大的通道达到几百cpm,远远超出期望限值1cpm标准,仪器已经不能正常使用。本论文通过系列排查、实验、论证,最终找出故障的根本原因,成功排除故障。并针对海盐地区夏季空气湿度大的特点,提出了控制优化措施,以确保LB770夏季稳定运行,避免此类问题的再次发生。
关键词
LB770;本底偏高;湿度;优化控制
中图分类号: TL817.2;R144 文献标识码: A
DOI:10.19694/j.cnki.issn2095-2457.2020.15.039
Abstract
With 10-channel α-β-Counter LB770Win regular calibration and maintenance,the background value was higher when doing the background measuring,it was focused on 1/2/6 channels,the largest channel reached several hundred cpm,far more than the expected limiting value of 1cpm,and the total β the PTR001BA sample which measured was only several cpm,the instrument was already not work well.This paper has found the root cause of instrument failure and successfully cleared the fault by series experiments and troubleshootings.In addition,this paper presents the series control and optimization measures about the characteristic of high humidity in summer in Haiyan county to make sure the stable operation of LB770 in summer,and avoid the happening problems of this sort.
Key Words
LB770;Background higher;Humidity;Optimization control
0 前言
压水堆核电厂,测量反应堆冷却剂中总α,以补充134I、固体裂变产生(140La、140Ba等)或239NP的测量,以确定是否存在燃料扩散的情况。在反应堆正常功率运行期间,如果没有燃料包壳破损,反应堆冷却剂中总α活度的期望值应小于0.01MBq/t,接近仪器探测限。通过对反应堆冷却剂中总β活度的测量,可以了解一回路的污染情况,以验证各种放射性测量的一致性。
LB770低本底α/β测定仪适用于对放射性样品的总α、β测量,它具有探头多、界面良好、可操作性强等特点。该仪器主要由10个流气式正比计数管、一个屏蔽计数管、前置放大器LB2025、铅室气体环路及计算机工作站等构成。工作气体使用氩保气(90%的氩和10%的CO2组成的混合气体),气流通过所有的计数管后,排出至户外。
依据《工器具维护保养》规程,每半年需要使用241Am和90Sr标准平面源分别对仪器测量总α和总β效率进行标定,对标定后的性能进行检测,以确保仪器处于最佳工作状态。2014年的一次定期標定维护中,在进行本底测量时发现1/2/6道的本底偏大,最大的通道已经达到几百cpm,远远超出期望限值1cpm标准,且测量的换料水箱样品总β仅为几个cpm的量级,仪器已经不能正常使用。……