高博跃 李辉 李明 张强
【摘 要】本文从实际生产出发,通过研究DGR和CGR伽马仪器的构造及原理,依据仪器特点利用自然伽马刻度原理,按照刻度误差要求,将CGR仪器的刻度装置运用于DGR仪器刻度.通过量值传递推导得出相应的API值,用于DGR仪器刻度,以便能在工作中替代非豁免的DGR刻度器,具有一定的经济价值及实用的意义。
【关键词】伽马;刻度;量值传递
中图分类号: P634.3 文献标识码: A 文章编号: 2095-2457(2018)14-0040-002
DOI:10.19694/j.cnki.issn2095-2457.2018.14.017
Gamma calibration device interchangeability test
GAO Bo-yue LI Hui LI Ming ZHANG Qiang
(China Oilfield Services Limited, Langfang Hebei 065201, China)
【Abstract】This article starts from the actual production, through the research of the structure and principle of DGR and CGR gamma instruments, using the principle of natural gamma scale according to the characteristics of the instrument. Through derivation of the value of the corresponding API value derived for the DGR instrument scale, in order to be able to replace non-exempt DGR cal in the work, has a certain economic value and practical significance.
【Key words】Gamma; Calibration; Value transfer
0 前言
γ射線与物质的相互作用,主要是通过光电效应、康普顿散射和电子对效应产生次级电子,这些次级电子能引起探测器内部离子的电离和激发。绝大多数仪器都是利用这两种现象探测γ射线。其中作为传感器的盖革米勒管与碘化钠晶体广泛应用于伽马测井仪器中。
1 DGR和CGR结构介绍
DGR仪器采用两组(每组8根)盖革米勒管,并呈180°对称布置,是SUB与INSERT结构,电路及探头放置在INSERT上;CGR仪器采用两根NaI晶体管,呈180°对称布置,其电路及探头都镶嵌在仪器的外壳中,并用盖板密封。
DGR和CGR仪器三级刻度都是采用钍刻度毯,其形状相同,只是API标称值与尺寸略有差别。其中在用的675DGR刻度毯API值为763.7API;675CGR刻度毯API值为200.5API。
如两种仪器的刻度毯可以互换使用,能在某些方面提高工作的效率。但是由于两种仪器的电路设计、仪器构造、金属材料及其厚度不同等原因,导致使用刻度毯互换刻度时,毯子上标称的API值有所不同。因此,需要通过量值传递实验进行研究,以确定相对应的API值,本文将使用CGR刻度毯对DGR仪器刻度,通过对刻度数据分析来获取相应的API值为例。……